Viscosity Bath & Vacuum Setup
- Product Code: IS:1206 (Part 2)
Equipment for testing Absolute Viscosity of Bitumen at 60 deg. C as per IS 1206 (Part 2) Rectangular model having double walled construction with toughened glass observation window (254 W x 225 H mm). Illumination from rear side provided. Oil immersion heater fitted to heat the bath media. Bath opening 254 x 200mm with cover having 4 openings of 50mm in diameter to hold 4 Asphalt Institute Vacu..
- Equipment for testing Absolute Viscosity of Bitumen at 60 deg. C as per IS 1206 (Part 2)
- Rectangular model having double walled construction with toughened glass observation window (254 W x 225 H mm). Illumination from rear side provided.
- Oil immersion heater fitted to heat the bath media.
- Bath opening 254 x 200mm with cover having 4 openings of 50mm in diameter to hold 4 Asphalt Institute Vacuum Viscometers with four viscometer holders.
- Bath liquid capacity 22 Litres.
- based PID temperature controller with two displays for set & process value with feather touch key board and PT-100 sensor & digital read out, resolution of 0.01°C having temperature accuracy of 0.02°C.
- Efficient stirring system.
- Temperature range :Above ambient to 70°C.
- Vacuum Set-up for Absolute Viscosity: Vacuum system capable of maintaining a vacuum to within ±0.05cm of desired level up to and including 30cm of mercury as per IS 1206(Part II) test method. Vacuum system comprises of vacuum gauge, receiver, valve and vacuum pump diaphragm type oil less. The whole system is mounted on a rectangular stand-alone plate with inlet and outlet.
- Power supply : 230 ±10% Volt AC, 1 Phase, 50 Hz.
Equipment Dimensions Shipping Dimensions Net Weight Shipping Weight
55W x 40D x 60H Cm 68W x 55D x 75H Cm 40.0 Kg 70.0 Kg